[. . . ] HI 99104 Warranty All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. [. . . ] If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. For technical service reference and for a complete list of available accessories, visit www. hannainst. com OPERATION · Connect the electrode to the meter and remove the black protective cap. If the electrode is dry, before taking measurements, immerse it in HI70300 storage solution for about 30 minutes to reactivate it. · Switch the meter ON, calibrate (see below), then immerse the electrode tip (4 cm/1. 5") into the sample. · After use, always turn the meter OFF, rinse the electrode with water, clean it with HI7061 cleaning solution and store it with a few drops of storage (HI70300) solution in the black protective cap. Note: When the display fades, replace the batteries by opening the compartment on the meter top. · Immerse the meter in pH7 buffer (HI7007), stir gently, allow the reading to stabilize and adjust the pH7 trimmer to read "7. 01" pH. · Rinse and immerse the electrode in pH4 (or pH10) buffer (HI7004 or HI7010), stir gently, allow the reading to stabilize and adjust the pH4/10 trimmer to read "4. 01" (or "10. 01") pH. [. . . ] · HI 70300M Storage solution, 230 mL bottle · HI 7061M Cleaning solution, 230 mL bottle S99104 06/04 [. . . ]